Calibration data for a Cambridge potentiometer instrument tested at 18 and 90 millivolt ranges.

Identifier  ExFiles\Box 165\3\  img050
Date  7th April 1924 guessed
  
X975 86033

CAMBRIDGE MEASURING INSTRUMENT
(POTENTIOMETER) (Mr. Hally).

18 millivolt range.
True reading from 22407. | Reading on Galvan. scale. | Ratio.
2.80 | 2.86 | .980 - 9
5.37 | 5.48 | .980 - 9
7.94 | 7.96 | .984 - 5
10.69 | 10.74 | .994 + 5
12.98 | 13.10 | .990 + 1
15.70 | 15.90 | .987 - 2
17.85 | 17.98 | .993 + 4

Average ratio .989
Subtract 1.1% from all readings, result then correct to within ±.8%
Scale proportionality V.fair.

90 millivolt range.
True reading from 22407. | Reading on Galvan. scale. | Ratio.
6.95 | 7.0 | .993 0
12.98 | 13.1 | .990 - 3
20.8 | 21.0 | .990 - 3
33.5 | 33.7 | .994 + 1
45.5 | 45.9 | .992 - 1
56.0 | 56.3 | .995 + 2
71.4 | 71.9 | .994 + 1

Average ratio .993
Subtract .7% from all readings, result then correct to within ±.3%
Scale proportionality good.
  
  