Rolls-Royce Archives
         « Prev  Box Series  Next »        

From the Rolls-Royce experimental archive: a quarter of a million communications from Rolls-Royce, 1906 to 1960's. Documents from the Sir Henry Royce Memorial Foundation (SHRMF).
Temperature measurement tests on two different types of coils (X and Y) under various conditions.

Identifier  ExFiles\Box 69\1\  scan0123
Date  6th July 1928
  
EFC3/T6.7.28. -2- Contd.

any air space round the coil, or rather between the coil and
the case. In this case one of the two coils, designated Y,
was of the finer primary type, and one of the coarser or
22 S.W.G. primary (now again standard) type, designated X.
The test consisted, as before, of :-

(1) Temperature measurements of the windings
by the increasing secondary resistance
method, after working five hours with
a 17.5 volt supply in conjunction with
a standard distributor set to the make
ratio of 345, running at 700 r.p.m.

(2) Similar temperature measurements after 1 1/2
hours standing on contact with a 12 volt
supply.

In each case a standard double ballast resistance
was used. We found the maximun temperatures attained to be :-

Coil. Condition 1. Condition 2.
Y 69°C. 65°C.
X 52°C. 40°C.

First comparing the temperatures attained by the
finer primary coil Y with temperatures previously attained
by these coils, in the same circumstances, the temperature
attained standing on contact is of the same order of magnitude.
The temperature attained after a five hours' run, however, may
be considered to be a little less than the average result in
the previous case. Further, as before, the temperature after
the run is slightly greater than the temperature after being
on contact. Now considering the result from the courser primary
coil X, the temperatures attained are considerably lower in

Contd.
  
  


Copyright Sustain 2025, All Rights Reserved.    whatever is rightly done, however humble, is noble
An unhandled error has occurred. Reload 🗙