The condition and electrical calibration of an Elliott Milliammeter No. 4808.

Identifier  ExFiles\Box 168\5\  img031
Date  22th September 1925
  
Y4622

EFC4/T22.9.25.

ELLIOTT MILLIAMMETER No. 4808.

This instrument appears to be in quite good order mechanically. It is nicely dead-beat and works without any sticking.

Electrically, the nominal range, without any external shunt, is .150 ampere (150 milliamperes, not 1500.)

The electrical calibration of this instrument is thus :-

Scale proportionality fair only.
Subtract 2.5% from readings above 125 divisions.
Subtract 3% from readings from 50 to 125 divisions and subtract 4% below 50 divisions.

The approximate resistance of the instrument between the terminals is .755 w at atmospheric temperature.

Millivolts for full scale deflection 110.


EFC.
  
  